Moisture meter common fault
We have 16 years' experience producing and exporting aluminum materials.
Trust in harmony,we care what our clients care.
Aluminum Pipe, Aluminum Tubing, China Aluminum Pipe, Aluminum Pipe supplier,heat pipe supplier,high frequency welding tube,micro channel multiport tube Trumony Aluminum Limited , https://www.szaluminumtube.com
Measurement time is too long. A: The closed mode that may be selected is not appropriate. Try replacing a suitable shutdown mode.
The instrument cannot be heated after starting the test. Answer: The overheating of the drying unit may cause the thermal overload protection function to start. For safe operation, the instrument is equipped with double overheating protection: the drying unit has a thermal overload protection (bimetal sensor) that shuts off the heating unit. In this case, contact your supplier.
Poor measurement repeatability. A: Possible reasons are: 1: The sample is not uniform. For example, they have different ingredients. The more uneven the sample, the more samples are needed to obtain better reproducible results. 2: The selected drying time is too short. Extend the drying time or choose the appropriate off mode "weight loss per unit time". 3: The sample is not completely dry (eg due to surface crusts). Try mixing quartz sand for drying. 4: The selected temperature is too high or the sample is oxidized. Reduce the drying temperature. 5: The sample boils and continuously splashes to change the temperature. Please clean the protective glass. 6: When the protection glass is eroded, the heating amount cannot be fully utilized. Please clean the protection glass. 7: The temperature sensor is contaminated or damaged. Please clean the temperature sensor. 8: The console is not stable. Please use a stable console. 9: The external environment is unstable (vibration, etc.).